Model : µpHCal50 | Make : ANALAB | HSN Code : 90278990
Supplied with Combination Glass pH Electrode (0-70ºC), Standard Buffer Tablets (7.00 pH, 4.01 pH & 9.18 pH), Temperature Probe, Electrode Stand consists of Clamp, Base Plate & Rod, Dust Cover, DC Adaptor, Operational & Instruction Manual with SOP, Calibration Test Certificate (without Traceability Document) and Warranty Card.
Technical Specification
pH Range : -2.000 to 20.000pH
Selectable Resolution : 0.1 / 0.01 / 0.001 pH
Relative Accuracy : ±0.02 / 0.002 pH
mV/ORP Range – mV / RmV : ±2000.0
ORP : EH auto calibration to standard hydrogen electrode
Resolution : 0.1
Relative Accuracy : ±0.2 mV OR 0.05%, whichever is greater.
Temperature Range : -5.0 to 130.0ºC
Resolution : 0.1ºC
Relative Accuracy : ±0.1ºC
Temperature Compensation : Auto / Manual
Temperature Calibration : In-built software and can be calibrated by users
Selectable Slope : 1) 80-120 %, 2) 90-110%
Total Numbers of Slope : 4 slopes for Five-point calibration & 2 slopes for Three-point calibration
Auto Buffer Recognition : 1) 4.01, 6.86 / 7.00, 9.18 / 10.01 pH @25ºC for Three points 2) 1.68, 4.01, 6.86 / 7.00, 9.18 / 10.01 & 12.45 pH @25ºC for Five points
Calibration Data Storage : Non Volatile Memory which can store, print & view calibration date & time, calibration done by, calibration point, pH, mV, temperature, % slope, validation pH for Calibration standard buffers, instrument serial number, instrument ID, model name etc.
Analysis Data Storage : More than 750 analysis data can be stored, printed & viewed with corresponding calibration data with company name, analysis date & time, record number, batch number / ID, sample name, username, pH, mV, temperature reading with corresponding calibration data, instrument serial number, instrument ID, model name etc. Secondly, analysis data can be print online without saving analysis data in instrument.
Searching of data : With record number, batch number, Date, Date Range and by scrolling.
Report Formation : As per GLP Compliance
Password Protection : Date & Time, Calibration Mode, Analysis Mode & Delete Stored Analysis Data.
Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port
Printer : 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)