Model : µMultiCal10 | Make : ANALAB | HSN Code : 90278990
Supplied with Combination Glass pH Electrode (0-70ºC), Conductivity Cell K=1.0, Standard Buffer Tablets, Temperature Probe, Electrode Stand consists of Clamp, Base Plate & Rod, Dust Cover, DC Adaptor, Operational & Instruction Manual with SOP, Calibration Test Certificate (Without Traceability Document) and Warranty Card.
Technical Specification
pH Range : -2.000 to 20.000 pH
Selectable Resolution : 0.1 / 0.01 / 0.001 pH
Relative Accuracy : ±0.02 / 0.002 pH
mV/ORP Range – mV/RmV : ±2000.0
ORP : EH auto calibration to standard hydrogen electrode
Resolution : 0.1
Relative Accuracy : ±0.2 mV or 0.05%, whichever is greater.
Slope : 80 - 120 %
Auto Buffer Recognition : 1) 6.86 / 7.00 & 4.01 / 9.18 / 10.01 pH @ 25ºC for two points
2) 4.01, 6.86 / 7.00 & 9.18 / 10.01 pH @ 25ºC for three points
Calibration Reminder Alarm : 0-99 hours
Result : Auto / Manual
pH Buffer Standard : US / NIST / EURO / DIN
Short Link : Provided to diagnose the problem either with Instrument OR Electrode.
Conductivity Range :
● 0-20.00, 20-200.0, 200-2000 µS/cm, 2-20.000, 20-200, 200-2000 mS/cm for K=1.0
● 0-10.00, 10-100.0, 100-1000 µS/cm, 1-10.000, 10-100, 100-1000 mS/cm for K=0.5
● 0-2.00, 2-20.0, 20-200, 200-2000 µS/cm, 2-20, 20-200 mS/cm for K=0.1
(Range value change depending upon cell selection)
Keypad : Membrane waterproof polyester soft keypad compliance to UL/ROHS.
Calibration Data Storage : Non-Volatile Memory can store, print & view calibration details.
Analysis Data Storage : More than 500 analysis data which can be viewed & Print, company name, analysis date & time, record number, *Username, *Sample Name, *Bath Number/ID, pH, mV, temperature reading, calibration data with date & time, conductivity at 25°C and conductivity at actual temperature, % Tempco, conductivity to TDS conversion factor, TDS, instrument serial number, Instrument Make & Model & *Signature etc. Secondly, analysis data can be print online without saving analysis data in Instrument.
(*: Not possible to enter data by instrument keypad but it will be handwritten)
Searching of data : With record number, batch number, date, date range and by scrolling.
Report Formation : As per Non-GLP Compliance
Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port
Printer : 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)