COMPARISON SHEET BETWEEN ALL MODELS OF ANALAB KARL FISCHER TITRATOR
Model |
µAquaCal5 |
µAquaCal10+ |
µAquaCal50 |
µAquaCal100 |
Specifications |
||||
Method of Detection |
Conductometric / Volumetric |
Conductometric / Volumetric |
Conductometric / Volumetric |
Conductometric / Volumetric |
Range of Moisture Detection |
Better than 100 ppm to 100% |
Better than 100 ppm to 100% |
Better than 50 ppm to 100% |
Better than 50 ppm to 100% |
Dispensing Resolution |
0.01ml |
0.01ml |
0.001ml |
0.001ml |
Magnetic Stirrer |
Built-in with soft touch Key speed controller |
Built-in with soft touch Key speed controller |
Built-in with soft touch Key speed controller |
Built-in with soft touch Key speed controller |
Calibration |
Average factor calculated
manually by |
Average factor calculated manually by |
Average factor calculated by |
Average factor
calculated by |
Method |
No |
No |
40 with Three pre-
assigned methods |
40 with Three pre-
assigned methods |
Auto Neutralization |
No |
On / Off - users selectable |
On / Off - users selectable |
On / Off - users selectable |
Stirring Time |
No |
00:00 to 99:00 (mm:ss) users adjustable |
00:00 to 99:00 (mm:ss) users adjustable |
00:00 to 99:00 (mm:ss) users adjustable |
End Point Sensitivity |
No |
80-100 - users adjustable |
80-100 - users adjustable |
80-100 - users adjustable |
End Point Time |
Fix 20 Seconds |
1-20 seconds - users adjustable |
1-20 seconds - users adjustable |
1-20 seconds - users adjustable |
End Point mV |
Fix 1000 mV |
1000-3000 mV – users adjustable |
1000-3000 mV – users adjustable |
1000-3000 mV – users adjustable |
End Point Detection |
By Audio & Visual Alarm |
By Audio & Visual Alarm |
By Audio & Visual Alarm |
By Audio & Visual Alarm |
Calibration Reminder Alarm |
No |
No |
0-99 Hours |
0-99 Hours |
Over titration Alarm |
Audio / Visual Alarm at 15 ml to save KF Reagent |
Audio / Visual Alarm at 15 ml to save KF Reagent |
Audio / Visual Alarm at 15 ml to save KF Reagent |
Audio / Visual Alarm at 15 ml to save KF Reagent |
Display of Reading |
Only mL of Reagent Consume |
Moisture in ppm, percentage (%), |
Moisture in ppm, percentage (%), mg/ml of water, gm/L, Mol/L |
Moisture in ppm, percentage (%), mg/ml of water, gm/L, Mol/L |
Auto Drain |
Yes |
Yes |
Yes |
Yes |
Flush |
Yes |
Yes |
Yes |
Yes |
Calculation of % RSD |
No |
No |
Yes |
Yes |
Average KF Factor Calculation |
Manually |
Manually |
Automatic – Maximum 5 factor stored in memory. |
Automatic – Maximum 5 factor stored in memory. |
Weight Entry |
NA |
NA |
In Resolution of 0.1 mg / 0.01 mg users selectable |
In Resolution of 0.1 mg / 0.01 mg users selectable |
Weight Transfer |
NA |
NA |
No |
through RS-232 Port |
Balance Interface |
NA |
NA |
No |
Yes, with RS-232 Port |
Dispenser Calibration/Validation |
Onsite by user |
Onsite by user |
Onsite by user |
Onsite by user |
Dispenser MOC |
SS-316L & Teflon |
SS-316L & Teflon |
SS-316L & Teflon |
SS-316L & Teflon |
Dispenser type |
Motorized |
Motorized |
Motorized |
Motorized |
Teflon Tube |
To transfer KF reagent from KF reagent bottle to dispenser & dispenser to reaction vessel |
To transfer KF reagent from KF reagent bottle to dispenser & dispenser to reaction vessel |
To transfer KF reagent from KF reagent bottle to dispenser & dispenser to reaction vessel |
To transfer KF reagent from KF reagent bottle to dispenser & dispenser to reaction vessel |
Material of Construction |
Epoxy Based Powder Coated Metallic Body / GMP Enclosure made up of Stainless Steel |
Epoxy Based Powder Coated Metallic Body / GMP Enclosure made up of Stainless Steel |
Epoxy Based Powder Coated Metallic Body / GMP Enclosure made up of Stainless Steel |
Epoxy Based Powder Coated Metallic Body / GMP Enclosure made up of Stainless Steel |
Type of Electrode |
Dual Platinum - Combined Type |
Dual Platinum - Combined Type |
Dual Platinum - Combined Type |
Dual Platinum - Combined Type |
Input |
Electrode with Banana Connector |
Electrode with Banana Connector |
Electrode with Banana Connector |
Electrode with Banana Connector |
Error Messages |
“Over KF”. |
“Over KF” & “KF Factor Error” &” Low Memory” |
“Over KF”, “KF Factor Error”, “Max. Factor Reached” & “Low Memory” |
“Over KF”, “KF Factor Error”, “Max. Factor Reached” & “Low Memory” |
Short Link |
Provided |
Provided |
Provided |
Provided |
Display |
16x2 Line Alphanumeric Backlit LCD |
16x2 Line Alphanumeric Backlit LCD |
20x4 Line Alphanumeric Backlit LCD |
20x4 Line Alphanumeric Backlit LCD |
Keypad |
Membrane waterproof polycarbonate soft touch keypad |
Membrane waterproof polycarbonate soft touch keypad |
Membrane waterproof polycarbonate soft t ouch keypad |
Membrane waterproof polycarbonate soft touch keypad |
Calibration Data Storage |
No |
Non-Volatile Memory |
Non-Volatile Memory |
Non-Volatile Memory |
Analysis Data Storage |
No |
< 500 |
< 750 |
< 1500 |
Searching of data |
No |
With record number scrolling. |
With record number, batch number/ID, date, date range and by scrolling. |
With record number, batch number/ID, date, date range and by scrolling. |
Report Formation |
No |
As per Non GLP Compliance |
As per GLP Compliance |
As per GLP Compliance |
21 CFR Part-11 Compliance |
No |
No |
No |
In – Built |
Password Protection |
No |
No |
Date & Time Calibration Mode Analysis Mode Delete Stored Analysis Data Delete Stored Analysis Method. |
For 40 users,10 Reviewers, 2 Admin & 1 ManufactureDiff ID and PW |
Printer Interface |
No |
Centronics Parallel Port / Serial (RS-232 C) Port |
Centronics Parallel Port / Serial (RS-232 C) Port |
Centronics Parallel Port / Serial (RS-232 C) Port |
Printer |
No |
40 / 80 Column Dot Matrix Printer |
40 / 80 Column Dot Matrix Printer |
40 / 80 Column Dot Matrix Printer |
PC Interface |
No |
No |
No |
USB Type – B |
PC Connectivity |
No |
No |
No |
By USB Interface Cable with Driver, ANALABsafe software to transfer (non-editable format) |
Power Supply |
230 VAC ±10%, 50 Hz |
230 VAC ±10%, 50 Hz |
230 VAC ±10%, 50 Hz |
230 VAC ±10%, 50 Hz |