Comparison Chart for Conductivity/TDS/Resistivity/Salinity/°C Analyzer

COMPARISON SHEET BETWEEN ALL MODELS OF ANALAB CONDUCTIVITY/TDS ANALYZER

Comparison Sheet

Model

ConCal+

µConCal5+

µConCal10+

µConCal50

µConCal100

µConCal

Handheld

Specifications

Conductivity Range 

0  to  200 mS/cm

0.1 to 200 mS/cm

(Distributed in Five Range)

0.1 to 200 mS/cm

(Distributed in Five Range)

0.1 to 2000 mS/cm

(Distributed in Six Range)

0.1 to 200 mS/cm

(Distributed in Five Range)

Resolution

0.001,0.01,0.1 µS/cm & 0.001, 0.01, 0.1, mS/cm

0.001,0.01,0.1 µS/cm & 0.001, 0.01, 0.1, mS/cm

0.001,0.01,0.1 µS/cm & 0.001, 0.01, 0.1, mS/cm

0.001,0.01,0.1 µS/cm & 0.001, 0.01, 0.1, 1 mS/cm

0.001,0.01,0.1 µS/cm & 0.001, 0.01, 0.1, mS/cm

TDS Range

0.1 to 100000 ppm  

0.1 to 100000 ppm  

0.1 to 100000 ppm  

0.1 to 1000000 ppm

0.1 to 100000 ppm  

TDS Resolution

0.001, 0.01, 0.1 ppm & 0.001, 0.01 ppt

0.001, 0.01, 0.1 ppm & 0.001, 0.01 ppt

0.001, 0.01, 0.1 ppm & 0.001, 0.01 ppt

0.001, 0.01, 0.1 ppm & 0.001, 0.01, 0.1 ppt

0.001, 0.01, 0.1 ppm & 0.001, 0.01 ppt

Resistivity Range

N.A.

N.A.

N.A.

 

20.00 MΩ to 50.00 KΩ, 5.00 KΩ, 500.00 Ω, 50.00 Ω, 0.50 Ω

N.A.

Resistivity Resolution

N.A.

N.A.

N.A.

0.001, 0.01, 0.1 KΩ & 0.001,0.01,0.1 MΩ

N.A

Salinity

N.A.

N.A.

N.A.

0.01 to 80000.0 ppm NaCl

N.A.

Temperature Range

0.0 to 130.0⁰C

0.0 to 130.0⁰C

-5.0 to 130.0°C

0.0 to 130.0⁰C

Resolution

0.1⁰C

Temperature

Compensation

Auto / Manual

Temperature Calibration

Built in Facility

Temperature Coefficient

0-10.000 % per ⁰C (Adjustable)

0-10.000 % per ⁰C (Adjustable) – One Per Range (total Five)

0-10.000 % per ⁰C (Adjustable) – One Per Range (total Six)

0-10.000 % per ⁰C (Adjustable) – One Per Range

Calibration Standard

One

Five Conductivity & TDS Standard - One Per Range

Five Conductivity & TDS Standard - One Per Range

Six Conductivity & TDS Standard - One Per Range

Five Conductivity & TDS Standard - One Per Range

Calibration Factor

One

Five – One Per Range

Six – One Per Range

Five – One Per Range

Range Selection

N.A.

Auto / Manual

Suitable Cell Constant (K)

1.0

0.1,0.5 & 1.0

1.0

Adjustable Cell Constant

0.90 to 1.1 K

0.09 to 0.11 K, 0.45 to 0.55 K & 0.90 to 1.1 K

0.90 to 1.1 K

Conductivity to TDS Conversion

0.40000 to 1.00000 (Adjustable Factor)

 Conductivity Display

@ Actual temperature and @ 25⁰C , TDS

@ Actual temperature and @ 25⁰C, TDS , Resistivity & Salinity

@ Actual temperature and @ 25⁰C , TDS

TDS Calibration & Analysis

No

No

In Separate Mode

No

Error Indication

“Wrong Standard / Cell”, “Cell Short”, “Try Upper Range”, “Try Lower Range”, “Temperature

probe Error”

Wrong Standard / Cell”, “Cell Short”, “Try Upper Range”, “Try Lower Range”, “Temperature probe Error” “Low Memory

“Wrong Standard / Cell”, “Cell Short”, “Try Upper Range”, “Try Lower Range”, “Temperature

probe Error”

Display

16 x 2 Line Alphanumeric Backlit LCD.

20 x 4 Line Alphanumeric Backlit LCD.

16 x 2 Line Alphanumeric Backlit LCD.

Keypad

Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS.

Calibration Data Storage

Non-Volatile Memory

Analysis Data Storage

No

No

< 500 data

< 750 data

< 1500 data

No

Searching of data

N.A.

N.A.

With record number by scrolling.

With record number, batch number/ID, date, date range and by scrolling.

N.A.

 Report Formation

No

No

Non GLP Compliance

As per GLP Compliance

No

21 CFR Part-11 Compliance

No

Yes – In-built with the system

No

Printer Interface

No

Centronics Parallel Port / Serial (RS-232 C) Port

No

Printer

No

40 / 80 Column Dot Matrix Printer

No

PC Interface

No

USB Type - B

No

PC Connectivity

No

USB Interface Cable with Driver with ANALABSAFE software

No

Power Supply

DC Adaptor

IQ/OQ/PQ Documentation

Yes - optionally

Calibration Certificate with traceability

Yes - optionally