Analab Scientific Instruments Pvt. Ltd. An ISO 9001 : 2000 Certified Co.
pH/mV/ºC/ORP Analyzer

pH / mV / ºC / ORP Analyzer
Model : µpHCal100



Request a Quote Now
Download Product Catalogue

Supplied with Combination Glass pH Electrode (0-70ºC), Standard Buffer Tablets, Temperature Probe, Electrode Stand with Clamp, Dust Cover, DC Adaptor, Operational & Instruction Manual, Manufacturer Test Certificate and Warranty Card.

Technical Specification :

  • pH Range : -1.000 to 15.000 pH
  • Selectable Resolution : 0.1 / 0.01 / 0.001 pH
  • Relative Accuracy : ± 0.002 pH
  • Range – mV / RmV : ± 2000.0
  • ORP : EH autocalibration to standard hydrogen electrode
  • Resolution : 0.1
  • Relative Accuracy : ± 0.2 mV or 0.05%, whichever is greater
  • Temperature Range : 0 to 130.0ºC
  • Temperature Resolution : 0.1ºC
  • Relative Accuracy : ± 0.1 °C
  • Backlit LCD Display : 20 x 2 line Alphanumeric
  • Auto Buffer Recognition : 1.68, 4.01, 6.86 / 7.00, 9.18 / 10.01 & 12.45 pH @25ºC
  • Buffer Standard : US / NIST & EURO
  • Power Supply : DC Adaptor

Salient Features :

  • Auto Buffer Recognition
  • Simultaneous Display of pH, mV & Temperature, Parallel & Serial Port for Printer & PC attachment
  • Report Formation as per GLP Compliance
  • More than 150 analysis data storage with corresponding Calibration data with Batch Number, Real Date & Time, Operator Name, Sample Name, searching of data as per Batch Number
  • Calibration Reminder Alarm (0-99 Hrs)
  • Results Hold / Un-Hold Facility etc.

Optional for μpHCal :

  • RS-232 to USB interface with Cable & Driver 100 Software to Transfer Data From Instrument to PC

Error Message :

  • Bad Buffer / Electrode Temperature Probe pH over range

Optional

  • NIST Traceable Standard for pH and ORP/Redox
  • IQ/OQ/PQ/DQ Documentation
  • Customized Electrodes
  • GMP Model on Request
  • Calibration Certificate with Tracability
  • Third Arm Stand


Analysis Report :

  • Company Name : ANALAB (VADODARA)
  • Analysis Date & Time : 13-NOV-2009 – 16:40
    pH = 04.291
    mV = +0160.2
    Temp. = 025.0

Calibration Report :

  • Company Name : ANALAB (Vadodara)
  • REPORT PRINT DATE & TIME : 13-NOV-2009 – 16:55
  • CALIBRATION POINT : FIVE
  • Calibration Date & Time : 19-Jan-2010 – 15.27
Std. Buffer 4.01
Std. Buffer 7.00/6.86   Std. Buffer 10.01/9.18
pH = 04.291 pH = 07.185 pH = 09.589
mV = +0160.2 Offset = -0011.0 mV = -0153.2
Temperature = 025.0 Temperature = 025.0 Temperature = 025.0
% Slope S1 = 096.77 Buffer used : 07.00 % Slope S2 = 110.27


Std. Buffer 1.68
Std. Buffer 12.45    
pH = 01.680 pH = 12.451  
mV = +0314.7 Offset = -0011.0  
Temperature =025.0 Temperature = 025.0  
% Slope S1: 100.09 % Slope S4 = 099.93  
 