Karl Fischer Titrator

Karl Fischer Titrator
Model : µAquaCal50




Request a Quote Now
Download Product Catalogue

Supplied with 250ml Reaction Vessel, Dual Platinum Electrode, Teflon Paddle Teflon Assembly , Moisture Trap for Reaction Vessel & 250ml Amber Reservoir Bottle, Dispensing Tube (02 Nos), Teflon Tubing for Dispensing Path, Operational & Instruction Manual, Manufacturer Test Certificate and Warranty Card.

Technical Specification :

  • Method of Detection : Conductometric
  • Backlit LCD Display : 20 x 4 Line Alphanumeric
  • Range of Moisture Detection : less than 20 PPM to 100%
  • Display of Reading : Direct Reading of Moisture in PPM, Percentage(%) & mg/ml of water
  • Dispenser : Motorized Dispensing for low and accurate measurement of reagent.
  • Electrode : Dual Platinum Electrode.
  • Stirring Time : 00:00 to 99:00 ( Minutes : Second ) - Users Adjustable
  • End Point Time : 1-99 Seconds , Users Adjustable
  • End Point Sensitivity of Moisture Detection : Users Adjustable
  • Calibration Reminder Alarm : 0-99 hours
  • Data Storage : More than 10 Analysis data with corresponding calibration data.
  • Interface : Parallel Port for Dot-matrix Printer attachment.
  • Input : Electrode Connector
  • Power supply : 230 VAC ±10%, 50 Hz

Salient Features :

  • Metalic Epoxy Based Powder Coated Body
  • Parallel Port to connect 80 column Dot-matrix Printer
  • Report Formation as per GLP Compliance
  • Built in Magnetic Stirrer with soft touch Key Speed Controller
  • Automatic end point detection with Audio Alarm
  • OVER TITER Audio Alarm save KF Reagent
  • Automatic Calculation of Avg. Factor and Maxi. 10 KF Factor can be stored in memory
  • Auto Neutralization reduces the time between analysis & Gives accurate Results
  • Onsite Dispenser Calibration/Validation Facility
  • MOC of Dispenser is SS/ABS/TEFLON for Corrosion Protection & Maintenance free Operation
  • In-Built Flushing System for cleaning dispensing path

Optional :

  • Auto Drain of Reaction Vessel
  • RS-232 to USB Interface with Cable & Driver software to Transfer Data from Inst. to PC
  • GMP model


Analysis Report :

  • Report Date & Time : 12-FEB-2010 – 16:09
  • Company Name: ANALAB (VADODARA)
  • Calibration Date & Time : 11-FEB-2010 – 23.34
  • Calibration Method : Water by Volume
  • Average Titer Factor : 6.666666
  • Analysis Date & Time : 12-FEB-2010 – 16.00
  • Sample Type & Method : Liquid by Volume
  • Volume of Sample (ml.) : 05.000
  • Density of Sample : 0.789
  • Stirring Time (Min:Sec) : 00:00
  • End Point Time (Seconds) : 20
  • End Point (ml.) : 01.41
  • Moisture %: 000.2382
  • Moisture (PPM) : 0002382.0
  • H2O (mg/ml) : 002797.87


Calibration Report :

  • Report Date & Time : 03-FEB-2010 – 17:04
  • Company Name : ANALAB (VADODARA)
  • Calibration Date & Time : 03-FEB-2010 – 15.22
  • Calibration Method : Water by Volume
  • Stirring Time (Seconds) : 00


Sr. No. Volume(microlitre) End Point(ml.) TF Value
01 10 01.99 5.025125
02 10 01.81 5.524862
03 10 01.81 5.524862
04 -N/A--    
05 10 01.81 5.524862
06 10 01.81 5.524862
07 --N/A--    
08 --N/A--    
09 --N/A--    
10 --N/A--    
Average Titer Factor 5.524862